Temperature | 1200℃ |
Voltage (AC) | 220V/380V |
Maximum Operating Temperature | 1250℃ |
Long-term Operating Temperature | 1200℃ |
Temperature Control Precision | ±1℃ |
Voltage | 380V/220V |
Power | 6KW |
Temperature Field Uniformity in Furnace | ±1℃ (depending on the size of the heating chamber); multi-point temperature control can be adopted for higher requirements |
Temperature Measuring Element and Measuring Range | Platinum-rhodium S, measuring range 0-1700℃ |
Number of Programming Curve Segments | Fully automatic heating, cooling and heat preservation; 50 segments for one group, 22 segments for two groups, 8 segments for three groups |
Communication | USB. The electric furnace can be controlled via computer for starting, pausing, stopping, heating, cooling, temperature curve setting, historical reports, etc. |
Heating Rate | Adjustable from 1℃/h to 40℃/min |
Heating Elements | Silicon carbide rods |
Installation Position and Method of Heating Elements | Vertically installed around the inner wall of the furnace chamber |
Refractory Materials | High-purity alumina zirconium-containing fiberboard |
Loading Capacity of Platform | 0.1-10 tons (customizable) |
Lifting Mechanism | Lead screw lifting |
Loading Platform In/Out | Manual/electric |
Lifting and In/Out Control | Fully controlled by PLC |
Cooling Structure | Double-layer furnace shell, air cooling + water cooling |
Water Cooling | Equipped with a circulating water pump and a circulating water tank (500 liters) |
Water-cooled Parts | Furnace door, furnace mouth, furnace top |
Vacuum Pump | Two-stage rotary vane pump |
Vacuum Value (Vacuum Gauge) | -0.1MPa |
Applicable Atmospheres | Hydrogen, nitrogen, argon, carbon monoxide, oxygen, etc. |
Gas Pressure Protection | To prevent dangers caused by the closure or blockage of the exhaust port in the furnace and excessive pressure in the furnace tube, this system is specially designed. Principle: Signals are obtained by an electric contact pressure gauge or pressure sensor, which drive the control module to close the electromagnetic intake valve and start the electromagnetic exhaust valve and alarm, so as to protect the normal operation of the electric furnace. |
Furnace Body Temperature | ≤45℃ |
Computer Interfaces | RS485, RS232, USB |
Warranty Scope and Period | The furnace is covered by a 1-year free warranty; heating elements are not covered. |