• +86-13526935931
  • gwdlthermo@gwdl.com

600°C Vacuum Aging Furnace Specifications

600℃ vacuum aging furnace, ±1℃ precise temperature control, 310S stainless steel furnace tube, three independent PID controls, 1.33×10⁻³Pa high vacuum, 30-segment program control.

600°C Vacuum Aging Furnace Specifications

600°C Vacuum Aging Furnace Specifications

Temperature:600
Vacuum Level:1.33×10⁻¹~1.33×10⁻³Pa (Pumping speed ≤30min)
Heating Elements:Resistance Wire Elements
Heating Rate:Adjustable (1°C/h to 15°C/min)
Tube Material:310S stainless steel (sandblasted)
Furnace Tube Dimensions (mm):Diameter 219 - Heating Zone 900 - Constant Temperature Zone 600

Product Video

Voltage AC380V
Rated Power12KW (Room temperature rises to 300°C within 30 minutes)
Long-term Working Temperature600°C
Temperature Control Accuracy±1°C
Furnace Tube MaterialStainless Steel Tube 310S (Sandblasted)
Inside TubeBuilt-in support plate (with holes)
Tube Load Capacity≤50kg
Tube Tray300mm (L) × 120mm (W) × 80mm (H), holds two pieces (2mm bottom holes, 3mm spacing), thickness 2mm
Furnace Tube Dimensions (mm)Diameter 219 - Heating Zone 900 - Constant Temperature Zone 600
Temperature Uniformity in FurnaceMeasured under no-load conditions in the constant temperature zone:600mm working zone: ≤±5°C 400mm working zone: ≤±4°C (Aging temperature 280-320°C) 
Thermocouple & Temperature RangeNickel-Chromium/Nickel-Silicon K-type, Range 0-900°C

Electrical Control System

Control Points3 independent thermocouples for temperature control, using Xiamen Yudian Intelligent Temperature Controller. Supports standard PID, AI-APID, or MPT tuning modes with self-tuning and self-learning functions. Features excellent control characteristics (no overshoot/undershoot), 10-100-segment program control, customizable ramp-up/cooling rates, and programmable commands (jump/loop, run, pause, stop). Allows real-time program modification during operation (temperature difference between three points ≤2°C).
Program Segments30 segments
Display & RecordingTouchscreen operation with automatic recording of temperature, time, and vacuum data
Compliant with Class II furnace requirements: temperature control indication ±0.2%, instrument accuracy 0.3%
Display ParametersTemperature, segment number, segment time, remaining time, output power percentage, voltage, current, etc.
Heating RateAdjustable from 1°C/h to 15°C/min
Heating ElementsHigh-temperature resistance wire (Nickel-Chromium alloy)
Heating Element PositionInstalled horizontally around the furnace tube
Refractory MaterialHigh-purity alumina fiber insulation board
Furnace Surface Temperature≤45°C

Vacuum System

Vacuum PumpMechanical pump + molecular pump unit
Vacuum Level1.33×10<sup>-1</sup>~1.33×10<sup>-3</sup> Pa (Pumping time ≤30min)
Vacuum GaugeDigital composite vacuum gauge
Flow MeterFloat flow meter (5L)
Process GasNitrogen
Gas InletLocated at bottom-left of furnace tube
Gas OutletLocated at top-right of furnace tube
Furnace Door MechanismHinged opening

Touchscreen Interface

Project NameParameters
LanguageChinese/English (switchable)
Full-screen Display & OperationInstrument screen, bar graph, historical trends, data reports, alarm information, data export, process flow, system management
Temperature ControlTouchscreen + high-precision integrated module
Start HeatingTouch operation
Pause HeatingTouch operation
Stop HeatingTouch operation
Temperature Curve ProgrammingTouch operation, 30 segments per curve
Real-time Status DisplayRunning curve name, segment number, segment time, elapsed segment time, digital temperature, real-time curve, power output percentage
Segment Selection StartTouch operation (cross-segment start)
Curve RangeAdjustable
Historical CurvesStorage for approximately 20 months
Data Reports (Excel)Storage for approximately 20 months, multi-point simultaneous display (USB expandable)
Data Logging IntervalAdjustable 1-3600 seconds, multi-point simultaneous display (USB expandable)
Alarm LanguageChinese display (alarm time and event description in Chinese)
Data Export InterfaceUSB (supports mouse)
Printer InterfaceParallel port
Touchscreen ProtectionPassword protection (operation prohibited without password)
Password ManagementAll touchscreen operations can be password protected to prevent parameter/process leakage and ensure comprehensive management

Warranty & Accessories

Warranty Coverage & Period1-year free warranty for electric furnace
Standard Accessories5 sealing rings, 1 instruction manual
Additional Remarks1.Nitrogen gas cabinet (capacity for 2 nitrogen tanks)
2.Nitrogen concentration analyzer
3.Audible & visual over-temperature alarm
4.Reserved temperature measurement ports (for 5 thermocouples)
5.Additional 5-point temperature measurement bracket (compliant with requirements)
6.Heating resistance wires (2 pcs)
7.USB flash drive (1 pc)
8.Feeding rod (1 pc)
9.Includes:
- Furnace temperature uniformity test system
- Vacuum gauge
- Thermocouples
- Temperature controller
- Nitrogen flow meter
- Nitrogen concentration analyzer

Additional Equipment

  • Lift door:Pneumatic door opening, controlled by foot pedal, with hot surface facing away from the operator.
  • Touchscreen: Multilingual menu, multiple temperature rise process curve storage, historical curve storage, and data reports (7 inches).
  • Computer Control: Controls furnace start, pause, stop, stores historical curves, sets temperature rise curves, reads temperature rise curves, sets parameters, etc.
  • Siemens PLC Control: Real-time data reading with open ports for data communication with upstream and downstream equipment.
  • Manual or automatic gas supply system
  • Observation Window: Designed on the furnace door with high-temperature-resistant quartz glass.
  • Movable Casters: Two swivel casters and two fixed casters.
  • Rapid Cooling: Achieved with a frequency inverter and centrifugal fan for programmable cooling control.
  • Quick Discharge Rack: The material rack is made of high-temperature steel, with wear-resistant heavy-duty refractory bricks on the top, enabling rapid
  • transfer of materials from the furnace to an external platform in high-temperature environments.
  • Multi-Point Temperature Control: Independent temperature control on multiple sides for improved temperature uniformity.
  • High temperature furnace chamber with multi-sided heating for very good temperature uniformity.
  • Stirring Fan: Metal stirring fan, can be linked with PLC to automatically start at designated temperature segments.
  • Independent Control Cabinet: Separate control cabinet with an extended control cable length (up to 5 meters).
  • Heat Storage Furnace Bottom: The furnace bottom sinks 50mm with a hollow design, supported by multiple vertical columns, increasing Ceramic Load-Bearing Plate and ensuring uniform heat distribution at the bottom.
  • Exhaust Gas Treatment: Adds VOC exhaust gas treatment furnace with a capacity of 40m³,H or 80m³,H.
  • Corrosion-Resistant Furnace Lining: Furnace lining material resistant to acids, alkalis, and hydrofluoric acid corrosion.
  • Heating Element Protection: Prevents contamination from powder and corrosion from volatile substances and scratch.
  • Clean Furnace Lining: Furnace lining made of ceramics, high-temperature glass, and high-temperature metals to prevent dust and volatile substances.
  • Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
  • Customizable charging racks according to requirements, for loading the furnace in two levels.

 

Connect with Us Today
Phone:+86-13526935931
Email:gwdlthermo@gwdl.com